12
Handling Telegram to Transmitter Telegram from Transmitter
Set correction factor to 4.2 Write request to transmitter "0021074206000420028
C
R"
of transmitter 2 Answer (O.K) "0021074206000420028
C
R"
Read current pressure value Read request to transmitter “0120074002=?108
C
R“
of transmitter 12 Answer represents 4.234 10
-5
mbar “0120074006423415040
C
R“
Adjust transmitter 1 at Write request to transmitter "0011074103001130
C
R"
atmosphere (high pressure) Answer (O.K) "0011074103001130
C
R"
Write request to transmitter "0011074006100023025
C
R"
Answer (O.K) "0011074006100023025
C
R"
# Name writable readable I1, I2 Date
040
1)
DeGas • • “01” “0”: abort respectively is not activ
“1”: activate respectively is activ
303 Error code • “06” “000000”: no error
“Err001”: transmitter defective
“Err002”: memory defective
312 Software version • “06” character string; six signs long
349 Transmitter type • “06” “.... A1” (32 32 32 32 65 49): CPT 100
“.... A2” (32 32 32 32 65 50): RPT 100
“.... A3” (32 32 32 32 65 51): PPT 100
“.... A4” (32 32 32 32 65 52): HPT 100
740 Current pressure value •
2)
• “06” “aaaabb” for a.aaa
··
10
(bb-20)
in mbar e.g. “123423” => 1.234 · 10
(23-20)
=>
1.234
··
10
3
=> 1234 mbar
741 Adjustment low/high • “03” “000”: low pressure
pressure “001”: high pressure
742 Correction factor • • “06” “abcdef” for abcd,ef
(Pirani) e.g. “000123” => 1.23
743
1)
Correction factor • • “06” “abcdef” for abcd,ef
(Bayard-Alpert) e.g. “000123” => 1.23
1)
only HPT 100
2)
see transmitter adjustment
DeGas (only for HPT 100)
The degassing mode can be activated/canceled with a write request and the status be
queried with a read request as well.
While the transmitter is in the degassing mode, no valid measuring values are transmitted.
Transmitter adjustment
To equalize a transmitter there must first be set a pressure adjustment point and then the
actual pressure value. If the setting for the pressure adjustment point is missing or if another
telegram is inserted between the two steps, an error signal is generated when writing the
pressure value.