16
8. Technical Data
8.1. Data List
Description Unit Value Comments
Measurement method Pirani / Bayard-Alpert
Material in contact with media Tungsten, Stainless Steel,
Nickel
Sealing Metal
Feedthrough Glass, Ceramics
Filaments Yttriated Iridium, double
Number of filaments 2
Measuring range Torr (mbar) 7.5 x 10
-10
... 750 (1 x 10
-9
... 1000)
Pressure Torr (bar) max. 4
Repeatability % of Torr (mbar) 7.5
x 10
-8
... 0.75 (1 x 10
-7
... 1) +/- 5 %
measurement
Accuracy % of 7.5
x 10
-8
... 0.75 (1 x 10
-7
... 1) +/- 10 %
measurement
Measuring cycles ms 40
Temperature: Operating: °C (°F) +5 ... +50 (+40 ... +120)
Temperature: Storage °C (°F) -40 ... +60 (-40 ... +140)
Temperature: Bake-out °C (°F) max. 356 without
electronics
Supply: Voltage VDC 24
Supply:
Power consumption max. W 7.5
Output digital RS 232 / RS 485
Serial interface 9600 baud, 8 databits,
no parity, 1 stopbit
Electrical connection 9 poles Sub-D socket,
screwable
Adjustment Pirani with display unit or software
to atmosphere and zero if pressure
< 3.75 • 10
-5
Torr ( < 5 • 10
-5
mbar)
Protection type IP 40/IP 54
Weight kg (lbs.) 0.4 (0.88)
Internal volume, approx. cm
3
1.0
Dimensions: KTR34130 mm (inch) 147 x 85 x 35 (5.8 x 3.4 x 1.4) incl. flange
KTR34330 mm (inch) 156 x 85 x 35 (6.2 x 3.4 x 1.4)
Flange: KTR34130 QF16 (DN 40 KF)
KTR34330 DN 40 CF
Measuring cable length m (ft) max. 1000 (3280), only for RS 485
Kommentare zu diesen Handbüchern