RM500SL
®
User's Guide Version 3.12
March 2015
1. First, complete an ANSI [Linear] or [AGC] test, then press <Tests>.
2. Highlight & <PICK> [Test box measures], then highlight and <PICK> [Telecoil].
3. Set the aid to ‘T’ (telecoil input).
4. Highlight and <PICK> [Start TMFS test]. Test setup instructions are displayed.
5. Position the aid on the ‘T’ inscribed on the test chamber floor. BTE aids should lie as flat as possible on the
test chamber surface, while custom aids should be oriented with their faceplate parallel to the top surface (if
possible) and touching it as shown.
6. Press <Continue> to start the test.
7. While maintaining the orientation of step 5, move the hearing instrument around on the test chamber surface
to maximize the number shown in the ‘Maximize Average-SPLITS Value’ box.
8. Without moving the hearing instrument, press <Continue> to store the Average (HFA-or SPA) SPLITS value
and run the SPLITS curve.
If an RTG has been previously obtained, the STS or RSETS will be shown.
Telecoil test results