07/02 ColorFlex 13-1
ColorFlex
Notice: Use of this equipment in a manner not specified by the manufacturer may
impair the protection afforded by the equipment.
The ColorFlex system consists of a ColorFlex optical sensor and an IBM-compatible computer
composed of the system unit, monitor, keyboard, and mouse. The computer may have been user-
provided or purchased with the sensor and software. A printer may also be used with the system.
The ColorFlex Sensor
The ColorFlex spectrocolorimeter is a versatile, compact color measurement instrument that can be used
on products of virtually any size and in industries as diverse as paint and textiles. It is available in two
different models, 45°/0° and diffuse/8° (sphere). The label on the bottom of the instrument identifies the
geometry. The instrument is designed for use with the sample port facing up or forward.
The Read Button
The Read (lightning bolt) button on the keypad, also known as the “Sensor button,” can be used to
perform any command or macro assigned to it through the Toolbar command on the Configure menu.
By default, this button activates a sample measurement. To assign it another function, see the Toolbar
section in the Configure chapter, Chapter 9.
LCD screen
(display)
Sample port
Keypad
Seitenansicht 0
1 2 3 4 5 6 ... 23 24

Inhaltsverzeichnis

Seite 1 - ColorFlex

07/02 ColorFlex 13-1ColorFlexNotice: Use of this equipment in a manner not specified by the manufacturer mayimpair the protection afforded by the

Seite 2 - The diffuse model

13-10 ColorFlex 07/02Calculations for the citrus scales are given in Appendix A.OJ4 Orange Juice StandardThis option provides an uncalibrated plastic

Seite 3 - Accessories

07/02 ColorFlex 13-11Skein Holder (45°/0° instruments only)This is a device for measuring yarn skeins. Wind the yarn around the skein holder in multi

Seite 4 - Options and Sample Devices

13-12 ColorFlex 07/02Foot SwitchPressing the foot switch performs the same function as the Read button onthe keypad. A pre-programmed sequence of ste

Seite 5 - 07/02 ColorFlex 13-5

07/02 ColorFlex 13-13InstallationThe following instructions guide you through the initial installation of your ColorFlex system.1. Unpack all cartons

Seite 6 - Bar Code Reader Kit

13-14 ColorFlex 07/02StandardizationStandardization on a ColorFlex model with 45°/0° geometry requires reading of the black glass andwhite tile. Stan

Seite 7 - 07/02 ColorFlex 13-7

07/02 ColorFlex 13-15Standardization can be done through the software or directly through the ColorFlex firmware. Moredetailed information about your

Seite 8 - Sample Cup Opaque Cover

13-16 ColorFlex 07/02MaintenanceThis section outlines the parts of the sensor you must maintain in order for the instrument to functionproperly. More

Seite 9 - 07/02 ColorFlex 13-9

07/02 ColorFlex 13-17Archiving SetupsAll ninety-nine setups can be stored in non-volatile memory by choosing “Archive Setups” from theColorFlex menu.W

Seite 10 - Tomato Red Tile

13-18 ColorFlex 07/02SpecificationsFor best performance, your instrument should be placed where there is ample work space with mediumor subdued illumi

Seite 11 - 07/02 ColorFlex 13-11

07/02 ColorFlex 13-19Baud rate from 300 to 57600No parity8 data bits1 stop bit.Your output device must be configured in the same way in order to commu

Seite 12 - Carrying Case

13-2 ColorFlex 07/02The diffuse model

Seite 13 - Installation

13-20 ColorFlex 07/02The instrument serial port ground connection is referenced to the instrument's internal frame and safetyground. Before conn

Seite 14 - Standardization

07/02 ColorFlex 13-217. When all readings are complete, examine the average DE* value. The instrument passes therepeatability test as long as the av

Seite 15 - 07/02 ColorFlex 13-15

13-22 ColorFlex 07/02DECLARATION OF CONFORMITYaccording to ISO/IEC Guide 22 and EN 45014Manufacturer's Name: Hunter Associates Laboratory, Inc.Ma

Seite 16 - Maintenance

07/02 ColorFlex 13-23Color Measurement NotesThe ColorFlex spectrocolorimeter is a versatile color measurement instrument that can be used onproducts o

Seite 18 - Specifications

07/02 ColorFlex 13-3AccessoriesThe following accessories are provided with the ColorFlex and can be found in the standards box.• Black glass - placed

Seite 19 - Power Required

13-4 ColorFlex 07/02Options and Sample DevicesAny or all of the following options and sample devices may be included. HunterLab part numbers areinclu

Seite 20 - Instrument Performance

07/02 ColorFlex 13-5Sample Clamp/Port-Forward StandThe port-forward stand holds the instrument securely in the port-forwardposition and the sample cla

Seite 21 - Regulatory Notice

13-6 ColorFlex 07/02Bar Code Reader KitThe bar code reader attaches to the communications port of theColorFlex and allows input of a 24-character bar

Seite 22 - DECLARATION OF CONFORMITY

07/02 ColorFlex 13-7To operate the bar code scanner, hold the wand at a slight angle (10 to 30 degrees) from vertical. Placethe wand on the page eith

Seite 23 - Color Measurement Notes

13-8 ColorFlex 07/02Note: If you do not use the autostore function, then be sure to manually log each reading bypressing the Up Arrow/Print key after

Seite 24 - 13-24 ColorFlex 07/02

07/02 ColorFlex 13-9opaque cover. Measure the sample at the sample port through the glass bottom of the sample cup withthe instrument in the port-up