Hp E1367A Bedienungsanleitung Seite 42

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 74
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 41
:COUNt? ARM:COUNt? [MIN | MAX] returns current number of scanning cycles set
by
ARM:COUNt. The current number of scan cycles is returned when MIN
or MAX is not supplied. With MIN or MAX passed as parameters, MIN
returns 1 and MAX returns 32767.
Parameters
Parameter Name Parameter Type Range of Values
MIN | MAX numeric MIN = 1, MAX = 32767
Comments Related Commands: INITiate[:IMMediate]
Example Query Number of Scanning Cycles
This example sets a switchbox for 10 scanning cycles of channels 00
through 03 and queries the number of scan cycles set. The
ARM:COUN?
command returns 10.
ARM:COUN 10 ! Set 10 scans per INIT command
ARM COUN? ! Query number of scans
42 HP E1366A/E1367A RF Multiplexers Command Reference Chapter 5
Seitenansicht 41